『Tech Support Questions 2』のカバーアート

Tech Support Questions 2

Tech Support Questions 2

無料で聴く

ポッドキャストの詳細を見る

Send us Fan Mail

The Inorganic Ventures team members answer more listener questions relevant to trace analysis. They address low mercury recovery at 2 ppb, noting mercury instability in nitric acid, adsorption to plastic, options such as preparing in HCl or stabilizing with ~1 ppm gold. They explain detection limit calculations using calibration data and blank replicates & distinguish instrument vs method detection limits, with suggestions to improve precision and sensitivity via conditions and sample introduction components. For heavy metals in blood studies, they warn of vacutainer/polypropylene contamination and recommend vessel leaching or vessel blank studies. Finally, they describe pH product certification against NIST standards, potential meter entry limitations, and temperature effects on calibration.

adbl_web_anon_alc_button_suppression_t1
まだレビューはありません